KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Creating and Editing a Scan Recipe
Automatic Parameter Adjustment
: - In general, a connection exists in the software
such that, when certain parameters are changed, other parameters are readjusted to
accommodate the changes. The adjustments occur between the
Y Scan Size
,
Traces
,
and
Y Spacing
parameters. After setting a parameter, the user might click on the other
parameters and notice a minor adjustment to the parameter that had just been set. This
happens to balance the numbers between
Y Scan Size
,
Traces
, and
Y Spacing
.
Show Position
- For 3D scans, the three options in this box are used for positioning the
scan area parameters box.
Table 3.15
Show Position
Options
Show Position
Option
Description
Graphic Representation
Start
The
Start
setting is used in the Video portion of the XY
view screen to position the upper left corner of the scan
area box in the Video screen crosshairs. Wherever the
user clicks in the video image, that position becomes the
upper left corner of the scan area box, and appears at the
center of the Video screen.
This is not the actual place where the scan starts.
Start
only defines the upper left corner of the scan
area box. Literal START is near the lower left corner.
Center
The
Center
setting is used in the Video portion of the XY
view screen to position the center of the scan area box in
the center of the Video screen crosshairs. Wherever the
user clicks in the video image, that position becomes the
center of the scan area box, and appears at the center of
the Video screen.
End
The
End
setting is used in the Video portion of the XY view
screen to position the lower right corner of the scan area
box in the center of the Video screen crosshairs. Wherever
the user clicks in the video image, that position becomes
the end of the scan area box, and appears at the center of
the Video screen.
This is not the actual place where the scan ends.
End
only defines the lower right corner of the scan area
box. Literal END is near the upper right corner.
Scan Feature
Click here to position
Outcome
Start
x
Scan Feature
Click here to position
Outcome
Center
x
Scan Feature
Click here to position
Outcome
End
x
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...