![KLA Tencor P-16+ Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647076.webp)
Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-28
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Changing the
Applied Force
setting:
1.
Click the menu arrow next to the variable box to display its menu.
2.
Click the desired force setting.
Stylus Radius (
μ
m)
Stylus Radius is the manufacturers stated radius of the stylus. The stylus radius cannot
be changed in this screen.
Use the Stylus Change Procedure to change the stylus radius setting. (See Stylus
Change Procedure on page 1.)
Recommended Maximum (mg)
Each stylus type is associated with a maximum applied force setting. The maximum
setting is deemed to be safe for the stylus and the sample while performing normal
scans. This force should not be exceeded.
Vertical Ranging Parameters (2D and 3D) - Scan Parameters Definition
3
Vertical Ranging
contains two parameters:
Range/Resolution
and
Profile Type
. These
two parameters are used together to set up the system for:
Range: The maximum feature measurement limit (theoretical), up or down, that
is considered when scanning for a feature,
Resolution: The theoretical vertical resolution of the electronics.
Three set of ranges are available. The primary differences between the ranges are in
their resolution capabilities, and the ability in the 131
μ
m, 327
μ
m, and 1000
μ
m
range to set the direction in which the range is applied. The ranges are described
below.
Table 3.16
Stylus Force Ranges for the Different Head Configurations
MicroHead V LF
MicroHead V SR
MicroHead V XR
0.05-50 mg
0.5-50 mg
0.5-50 mg
CAUTION:
Recommendations and limits are only correct if the “Stylus
Change Procedure” was followed when the stylus was installed.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...