0142530-000 AB
KLA-Tencor Confidential
6-27
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Sequence Recipe and Data - Editing the Options Field in the Sequence Editor
Teaching the Base Angle (P-16+ Only)
6
The Base Angle is an offset angle relative to the orientation of the sample’s pattern. It
is used to align scans with the wafer geometry. It is to be used primarily for scan
sequences using manual load in conjunction with the
No Automatic Load/Unload
handler option in the system. (The Base Angle is fixed for all scans in the sequence.
Use the following procedure to program the Base Angle.
1.
Double-click
Base Angle
in the sequence list. (See
2.
A warning is displayed before the Teach Location screen appears. The warning
says that deskew and measurement sites could be invalidated. Click
OK
to
proceed or
Cancel
to abort the procedure.
3.
Before the Teach Location screen appears, a warning message is generated,
stating that updating the Base Angle may invalidate the Deskew sites.
4.
The Teach Location window appears. Locate a line or other pattern to use for a
reference.
5.
Click the clockwise or counterclockwise
Rotation
buttons in the toolbar until the
crosshair is aligned with the reference feature.
Figure 6.26
Sequence Editor
Double-click
Base
Angle
to open the Teach
Location screen.
NOTE:
As the range rotates, if necessary, move the stage to keep the
reference feature in the field of view.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...