Sequence Recipe and Data - Using Multi Analysis In Sequence
KLA-Tencor P-16+ / P-6 User’s Guide
6-34
KLA-Tencor Confidential
0142530-000 AB
3/13/09
3.
Click the
Recover
button.
Calculating Combined Sequence Statistics
6
Values from different sequence sets can be combined into one, and used to calculate
the standard deviation, mean, and so forth. The computer accesses stored data from
selected data sets in the Sequence Data catalog and recalculates them.
1.
Click the
Sequence Data
command button in the Catalog screen.
The Sequence Data catalog window appears.
2.
Highlight the data files to be combined.
3.
Press
CTRL
while clicking to highlight multiple data files.
4.
Click the
Combine
button.
5.
Enter a name for the new combined data set.
6.
Click
OK
.
A statistics summary with the new data appears after a short calculation interval.
U
SING
M
ULTI
A
NALYSIS
I
N
S
EQUENCE
6
Multiple data analyses can be obtained from a single scan by applying the data
analysis settings of additional recipes to its raw data. The process is a modification of
a sequence recipe in which the instrument uses the first scan recipe to scan and
analyze in the usual manner, then takes settings from the subsequent recipes to
reanalyze the first recipe’s scan data.
It is important to note that the raw data for the scan be saved and therefore can be
subjected to numerous different parameter adjustments. Each set of data that is
obtained from applying the new parameters can be save under its own name. This
means that after the scan is run and the results saved, the additional information can
be retrieved at a later date, even calculated on a desktop version of the software if it
has been purchased.
Time can be saved and throughput improved by using multiple analysis for:
Measurements that require more than one cursor setting — such as two different
step heights on a single scan
Measurements with different filter settings
Measurements with different surface parameters enabled in the Scan recipe.
1.
Go to the
Sequence Recipe
catalog window, and select a Sequence recipe.
2.
Click the
View/Modify
button to open the Sequence Editor window.
3.
Click the
New
button at the bottom of the screen or click the
Sequence
menu,
and select
New
.
A blank sequence list appears.
4.
Set up the scanning recipe to scan with its existing settings:
a.
Click the name of the required recipe to be used for the scan.
b.
Click
Add
to add the Scan recipe to the list.
5.
To make changes to an existing Scan recipe.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...