Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-64
KLA-Tencor Confidential
0142530-000 AB
3/13/09
1.
The option exists to create three 2D bearing ratio parameters. Click in up to
three empty checkboxes to put a check (
9
) in them and activate their variable
2.
The depth is set down from the highest peak in the scan. It can be set in either
microns (
μ
m) or angstroms (Å). Determine which units are going to be used and,
if a change is necessary, click the menu arrow under units to display its menu.
3.
Click the desired unit signifier,
μ
m
or
Å
.
4.
Depth
is the distance down from the top of the highest point in the scan. To set or
change the
Depth
, double-click the current
Depth
variable and type in the new
depth. (See
2D Cutting Depth (CutDp)
3
Cutting Depth
is related to Bearing Ratio in that Bearing Ratio uses an operator set
depth from the top peak in the scan, adding up the points between the top peak and the
set depth, while
Cutting Depth
uses an operator set
ratio of data points
in the scan that
are below the highest peak in the scan, causing the system to determine the depth.
(See the definition of
Bearing Length Ratio
Use the following procedure to set the 2D Cutting Depth variables.
Figure 3.48
2D Bearing Ratio
Figure 3.49
Depth for 2D Bearing Ratio
L
(Sampling Length)
S
2
S
1
Depth
Highest point
in the scan.
Depth: Set a number of
μ
m or Å down from the
highest point.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...