0142530-000 AB
KLA-Tencor Confidential
6-9
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Sequence Recipe and Data - Creating a Sequence Recipe
12.
The
Pattern Rec. Deskew Teach: Site 1
screen is displayed. Click a die in the
upper left quadrant of the sample navigation grid. the dark (blue on the screen)
rectangle has been chosen in
13.
After the die in the upper left quadrant is clicked, the system moves that die into
view in the video window. Click
FOCUS
in the tool bar to bring the die into clear
focus.
14.
Use the arrow buttons in the tool bar to move the field of vision to a feature in
that die that is used for centering the die and aligning the wafer. It is best to use
the same feature that is used in the die grid. (See
.)
15.
After locating the feature, use the click and drag procedure, starting from the
upper left corner of the feature, to draw a rectangle around the feature. When the
box is complete, the system centers it in the X-Y-grid and a replica of it is
produced in a box on the screen. The die navigation box, under the die grid
navigation grid, now contains a small blue box indicating the position of the
feature with respect to the die boundaries. (See
16.
When satisfied with the die position and the feature, click
OK
.
End
: Teach Global Pattern
Recognition Sites
17.
through
for
Site 2
. For a location, choose the lower right
quadrant, at approximately the opposite die position, at an approximate 45°
angle through the center of the die grid from the first die.
Figure 6.10
Pattern Rec. Deskew Teach: Site 1 Screen
(The video image is not
visible in this illustration but
would be present on the
actual screen.) The box is
drawn around the feature
that is displayed in the
Current Patt...
box.
Click a
die in the upper
left quadrant.
The location of the
feature is indicated
in the die navigation
box.
Use the
click-and-drag process to
draw a box around the
feature to be used for
pattern recognition deskew.
Click
OK
when satisfied with
the feature and die
location.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...