![KLA Tencor P-16+ Скачать руководство пользователя страница 242](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647242.webp)
Sequence Recipe and Data - Groping with Pattern Recognition (P-16+ Only)
KLA-Tencor P-16+ / P-6 User’s Guide
6-46
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Lowest Match Score
(Was changed from
Minimum Match
Score, which is still
the term used in the
Configuration
screen version of
this parameter.)
Lowest Match Score is used to compare all the groping
positions in the given groping levels. Once the groping stops
(assuming that the Minimum Score to Stop Groping is not
found) the highest score achieved, among those scores that
qualified for Lowest Match Score acceptance, is chosen as
the search pattern (model). This number must be smaller than
the Minimum Score to Stop Groping. Allowed values range
between 20 to 100%; the default is 65%.
This parameter allows adjustment of the threshold at which
the pattern recognition system concludes that it has found a
candidate for the desired deskew site.
Minimum Score To
Stop Groping
Minimum Score to Stop Groping defines a value at which the
system accepts the image as the model for which it is
searching. Groping stops if this score is reached and the
image corresponding to the score is considered to be the
search pattern (model).
EXPLANATION: If the pattern recognition system is groping
to find the desired pattern, frequently the matching pattern is
found with little ambiguity. If a score equal to or better than
the Minimum Score to Stop Groping occurs, the searching
process stops and the deskew site is placed. Allowed values
range between 20 to 100%; the default is 70%.
If no matches are found that are as good as this setting, the
search continues until all retry layer areas are searched. The
highest score above the Lowest Match Score setting
determines the placement of the deskew site.
Edge Based Pattern
Recognition
The
Edge Based Pattern Recognition
option is used for low
contrast image recognition on a sample surface or where
there is a large surface light variation. If this option is chosen
(with a check in the check box), the normal image contrast
grayscale processing takes place first, then a series of filters
are applied that further contrast and sharpen edges for a
better pattern recognition.
The image data is stored before these filters are applied so
the data is not effected by this option. It is strictly a tool used
for pattern recognition where contrast is low or where light
varies significantly.
If the option is not chosen, only the image contrast grayscale
processing is performed..
Save/Apply Video
Settings
The lamp brightness setting is important in pattern
recognition. If the lamp brightness is different from when the
original sequence was established, the pattern recognition
images could be difficult for the system to detect. A check in
the
Save/Apply Video Settings
checkbox ensures that the
lamp brightness is saved with each deskew site pattern so
future scans have the same image view with the same light
for pattern recognition.
Table 6.12
Groping Parameters (Continued)
Parameter
Description
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...