![KLA Tencor P-16+ Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647068.webp)
Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-20
KLA-Tencor Confidential
0142530-000 AB
3/13/09
1.
Multi-Scan Average
- This is a 2D option that allows the user to repeat a single
scan up to 10 times so that the scan data can be averaged by the number of scans
performed. This feature provides an opportunity to minimize the noise factors in
a scan.
2.
Scan Direction - Arrow
- This option dictates the direction of the scan, from left
to right
or from right to left
.
ye Changing The Scan Direction
: Click the arrow to cause it to point the
opposite direction.
3.
Scan Direction - Teach
- When the
Teach
button is clicked on, it opens in the XY
View. This screen allows the user to set the scan position and scan size.
a.
Click the radio button next to the desired reference position,
Start, Center
,
or
End
, that is to be established with respect to the scan feature in the
Teach
Scan Length
screen. This can also be changed in the XY View to allow the
user to view the start, middle, and end of the scan, which can be very
useful when the scan size exceeds the field of view.
NOTE:
DO NOT use
unless it is absolutely necessary.
The recommended direction is left to right
because it
gives better repeatability, protects the stylus, and provides better data.
NOTE:
The Reference Position is for user convenience only. The
coordinates stored in the sequence recipe are the always scan center.
This allows the user to change the scan size and have it expand or
contract about the center of the feature.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...