recipe, the number of Traces is limited to 2000. When setting up a sensor
stage recipe, the allowable number of Traces is limited to 200 and is also
limited by the recipe y size, and Feature Find Y Spacing and any offsets, with
the total summing to less than the 90 µm scan size limitation for the sensor
stage. In addition, if Oval / Rectangle search type is used, the Y Size for this
scan also limits the number of traces.
Y Spacing (µm)
3
Y Spacing is the distance between traces. When searching for an isolated
feature, the Y Spacing should be set at about two thirds of the feature size. As
shown in a previous example, when searching for a 5 µm box, the Y Spacing
should be set to about 3.5 µm. If it is set larger, then the Feature Find scan can
step over the box entirely and not find it during the scan. If the Y Spacing is
set smaller than 3.5 µm, then throughput is decreased since a larger number of
lines are required to find the feature. When searching for a feature in an array
of identical features, the Y Spacing should be set at about two thirds of the
pitch between features in the y direction. Again, this allows for a quicker
search than setting at a smaller value. It should never be set at the pitch in the
y direction since this can result in no features being found since each x scan
line will be run on similar features.
Scan Speed (µm/s)
3
The Scan Speed should be set with the same guidelines as discussed in the
Contact Mode section of the scan recipe chapter.
Sampling (Hz)
3
The Sampling Rate should be set with the same guidelines as discussed in the
Contact Mode section of the scan recipe chapter.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...