0142530-000 AB
KLA-Tencor Confidential
3-95
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Feature Find
filter size for the Feature Find scan. So if the tolerance is set too large, then a
large filter can be applied which will affect the scan and Feature Find search
performance. In addition, if there are similar features that would be contained
in the Feature Find scan at a different width, setting it too loose can result in
finding the wrong feature, similar to the Feature Depth / Height restrictions.
Model
3
shows the Feature Find definition for Model Feature, which was
discussed earlier in the chapter. Model Feature does not have the user define
the dimensions of the feature in x, y, and z, but instead uses scan data as a
model and compares this model to the Feature Find scan data as it is collected
in a pattern recognition type operation. For Model Feature the user loads a
model, which is a 3D scan data file. The model is created by running a 3D
scan or by cropping a portion of a 3D scan to form the model. Cropping is
discussed in more detail in a separate chapter of the User's Guide.
The model used must contain a minimum of three lines. When setting the X
Scan Size in Feature Find definition, the scan length must be larger than the
model size, with three times the model size the minimum scan length
recommended. An even larger Feature Find scan is recommended to account
for positioning error. In addition, the Y Spacing should be set at the same
value as the Y Spacing from the model. Although Feature Find will
interpolate the model Y Spacing if it does not match Feature Find Y Spacing,
this results in unpredictable pattern recognition performance, so it is not
recommended. The number of Traces must be set at a minimum equal to the
number of lines from the model, but a larger number is recommended to
account for positioning error.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...