![KLA Tencor P-16+ Скачать руководство пользователя страница 115](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647115.webp)
0142530-000 AB
KLA-Tencor Confidential
3-67
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Creating and Editing a Scan Recipe
High Spot Count and Peak Count
3
Access the High Spot Count and Peak Count Display Window by clicking the
High
Spot Count/Peak Count
button in the Recipe Editor. (See
.)
High Spot Count (HSC)
3
High Spot Count is the number of profile peaks per unit of length projecting through a
reference line parallel to and at a given height above, a line drawn parallel to the mean
line through the lowest point of the roughness trace. (See
).
The mean line is the line at the mean height of all data. Another line is drawn through
the lowest point in the trace, parallel to the mean line. The reference line is at a user
specified height above the lowest point line.
Figure 3.51
Bearing Ratio and Cutting Depth Parameters
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...