Sequence Recipe and Data - Correlation Scans
C
ORRELATION
S
CANS
6
Scans are correlated when a long scan is performed first, then small scans are
performed in the same general location. Correlation scanning combines local area
scans with macroscopic scans so that discrete features can be related to global surface
planarity.
From the scan data of a long scan, distinct features can be located which require a
repeat scan at high resolution, then create a sequence that performs high resolution
sub-scans along the length of the long scan. Data for each sub-scan is based on the
long scan, providing a data reference for correlating the measurements of the
sub-scans.
1.
Open an existing Sequence recipe or create a new one in the
Sequence Editor
.
2.
Select the recipe to use for the long scan (one that traverses the targeted feature).
3.
Click the
Scan Type
arrow below the sequence to open the list.
4.
Click the
Correlation Long Scan
button.
A message dialog box appears, warning that the recipe immediately following is
designated a Correlation Sub-scan and if it is set up for multiple analysis, it
resets to single scan. The Sub-scan is the short scan that is tied to the long scan.
It provides the local, small-scale analysis. It is set up in step 7, next page.
5.
Click
OK
.
The long scan recipe becomes red; the recipe immediately following becomes
blue, indicating that it is a sub-scan to that long scan. Sub-scans always follow
long scans in sequence.
6.
Designate the other sub-scans (usually 100 µm or less) as done for the long scan,
using the
Scan Type
list to select
Correlation Sub-scan
.
shows the Sequence Editor for a correlation scan where the parent
long scan recipe is
EXAMPLE
, Loc is location 1. The sub-scans are
EXAMPLE2
,
Loc are location 2, location 3, and location 4.
NOTE:
Multiple correlation scan sets can be established; each set is
marked by the initial long scan in red lettering and its associated
sub-scans immediately following in blue.
Figure 6.29
Correlation Scan Sequence
The first recipe, the parent
long scan, is in red.
The subscan recipes, the
correlation subscans, are
in blue.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...