The system is facilitated by a series of interconnected calibrations. The
interdependency of the calibrations makes it important that those who calibrate the
systems understand the which calibrations affect other calibrations. When performing
any of the calibrations for the system, ensure that all prerequisite calibrations are
performed prior to performing the target calibration. When the target calibration is
completed, ensure that any necessary subsequent calibrations are performed or the
possibility exists for inaccurate scans.
A
PPLIED
F
ORCE
C
ALIBRATION
12
Check the Calibration
for
possible interaction with
other calibrations.
Introduction
12
Applied force is the force between the stylus tip and the sample when the stylus is in
contact with the sample. Mechanical changes in the stylus arm can affect calibration
settings.
Table 12.1
Standard Calibration Matrix
Calibration to be
Performed
Calibration
Prerequisites
Post Calibration
Requirements
System Performance Results
Applied Force
none
none
Protects stylus and sample during nulling procedure.
Video Calibration
Applied Force
none
Objects chosen (clicked on) in the screen are
accurately positioned in the center of the screen.
Improves accuracy of pattern recognition deskew and
site-by-site pattern recognition.
Scan Position
Offset Calibration
Applied Force,
Video
none
When performing a scan with the sample stage, the
general location taught for the scan is accurate. The
scan occurs very near the taught position.
Step Height
Applied Force,
Scan Position
Offset
none
Feature steps on the sample surface are more
accurately measured.
Radius of Curvature Applied Force,
Step Height
none
Radii of curved surfaces are more accurately
measured.
Level
Applied Force
none
Ensures that the stylus does not exceed its vertical
range due to the excessive tilt or level orientation of the
stage.
Lamp Balance
Applied Force,
Drop Timer
none
Drop Timer
Applied force
none
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...