PFX2532
137
Troubleshooting
App
.
is displayed on the
front panel.
An alarm occurred on a connected electronic load. Are the electronic
load’s protection functions (e.g., OPP) set properly? Is the electronic
load turned on?
is displayed on the
front panel.
Are the external analog control settings of the connected electronic
load correct? Are the flat cables connected properly? Is there a
problem with connection cables or DUT (battery)?
This alarm may occur if discharging causes the rated power of the
electronic load to be exceeded.
is displayed on the
front panel.
An alarm occurred on a connected DC power supply. Are the DC power
supply’s protection functions (e.g., OVP, OCP) set properly?
Is the DC power supply turned on?
is displayed on the
front panel.
Are the external analog control settings of the connected DC power
supply correct? Are the flat cables connected properly?
Is there a problem with connection cables or DUT (battery)?
This alarm may occur if charging causes the rated power of the DC
power supply to be exceeded.
is displayed on the
front panel.
The initial current (load check current that is smaller than the set
current) for checking the DUT status is not flowing when attempting to
start a charge, pattern or IV characteristics test.
Is the DUT connected correctly? Is the charging circuit cut off by the
battery protection circuit (BMS)?
is displayed on the
front panel.
Are the LAN and flat cables connected properly?
is displayed on the
front panel.
A system with the same IP address was detected on the LAN
communication circuit. When using fixed IP addresses, assign unique IP
addresses using IO Config of BPChecker3000. If you are using several
PFX2532s, be sure also to set the channel numbers.
An I/O (Comm) alarm is
displayed on the application
window.
This appears when communication between the PC and the PFX2532
is not working. Check the system connections. This alarm also occurs if
the PFX2532 power supply is cut off temporarily. For details on how to
restore the system after a power outage, see the BPChecker3000 help
file.
Test Executive
Symptom
Items to check and possible remedy
See
?
Help
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......