50
PFX2532
Configuring the Electronic Load
Setting Protection Functions
Configure the protection function.
If the Electronic load OCP or other alarm is activated during a charge / discharge test, CD/B alarm
occurs on the PFX2532, and the load is turned off. Press ENTER on the Electronic load, and eliminate
the cause of the alarm. Then, clear the alarm from BPChecker3000.
For details on the setup procedure, a description of the alarm that occurs when a protection
function is activated, and how to clear the alarm, see the operation manual of the Electronic load
that you are using.
Setting the Slew Rate
Configure the slew rate.
Configuring External Analog Control
■
Configuring the PLZ-5W
1
Use the Menu key.
• CC/ CR/ CC control
SOURCE > MORE > External Control> CC/ CR/ CC > Edit
:
Enable
• Load on/ off control
SOURCE > MORE > External Control > Input Logic > Edit
:
Negative
PLZ-5W
Set to 110 % of the maximum discharge current for overcurrent protection (OCP).
PLZ-4W
Set to 105 % of the maximum discharge current for overcurrent protection (OCP).
See
PLZ-5W
Press SOURCE > Level >Slew Rate, and then use the numeric keypad or the rotary knob to set
the maximum value.
PLZ-4W
Press SLEW RATE, and then use the rotary knob to set the maximum value.
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......