Configuring the Electronic Load
To protect the connected devices, set the protection function of the electronic load.
Configure external analog control to enable the electronic load to be controlled from the PFX2532.
Set the number of electronic loads and boosters.
PLZ-5W, PLZ-4W
After configuration, we recommend that you lock the panel. This will prevent the settings from
being changed accidentally.
Electronic load configuration procedure
1
Initialize the Electronic load to its factory default settings.
The contents in memory will also be cleared.
2
Configure the protection function.
Set the overcurrent protection (OCP).
3
Set the slew rate.
Set the slew rate value.
4
Configure external analog control.
Set the external voltage control, response, load on / off logic, and EXT CONT.
5
When connecting PLZ-4Ws in parallel, configure the master unit and slave
unit. When connecting a booster, set the number of boosters.
6
Enable the panel lock.
Set the lock function.
Ope
PLZ-5W
:
Press SYSTEM > Admin > Sanitize > Execute to show a confirmation
screen. Select OK, and press ENTER
PLZ-4W
:
While holding down ENTER, turn the POWER switch on.
See
See
See
See
See
• Even when the panel is locked, the load on / off operation is not locked. Note that if you turn on
or off the load during a charge / discharge test, it could destroy the test data.
• No adjustment is necessary between connected devices.
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......