108
PFX2532
Preparing to Connect the DUT (Battery)
Cables
1
for connecting the OP03-PFX and the input terminal to the DUT (battery) are not included.
Prepare the following cables.
■
Cables and tools that are necessary to make the connection
1
The TL11-PFX / TL12-PFX [CE compliant product] option consists of a dedicated voltage
sensing cable and thermocouple for connecting the DUT (battery) to a PFX2532 in which an
OP03-PFX is installed. (
See
Wire
DUT cable
Cable for running charge/discharge current between the PFX2532
and DUT (battery)
Voltage sensing cable
Cable for sensing the electrode voltage of the DUT (battery)
Cell voltage sensing
cable
Cable for sensing the cell electrode voltage of the DUT (battery)
0.20 mm
2
(AWG24) to 0.52 mm
2
(AWG20), shielded cable
Flat-blade screwdriver
Shaft diameter: Ø3, blade tip width: 2.6 mm
Wire stripper
A wire stripper that matches the wires listed above
See
+
–
+
–
+
–
+
–
+
–
+
–
+
–
+
–
OP03-PFX
Cell voltage sensing cable (shielded cable)
Cell V
Term1
Cell V
Term2
Cell V
Term3
Cell V
Term4
Cell V
Term5
Cell V
Term6
Cell V
Term7
Cell V
Term8
SL01-PFX
The DUT (battery)
Example of connecting the DUT (battery) to an SL01-PFX after installing an Voltmeter Unit OP03-PFX
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......