44
PFX2532
Connecting the DUT (Battery)
Connecting the PFX2532 to the DUT (Battery)
1
Check that the PFX2532 is ready to execute tests (power-on state).
You can connect or disconnect DUT cables and sensing connectors even when the
power switch is on. But, make sure that a charge/discharge test is not in progress.
2
Connect DUT cables to the positive and negative DUT terminals of the
PFX2532.
Connect the polarities according to the colors that you decided to use when creating
the cables.
Run the cables out as horizontal as possible, and fix them in place with screws.
3
Connect the other end of the DUT cables to the DUT (battery).
Connecting sensing cables
1
Insert the assembled sensing connector into the PFX2532 SENSING connector.
Secure it in place by fastening the connector cover screws.
2
Connect or attach a voltage sensing cable and thermistor to the DUT
(battery).
Screw (M8)
DUT cable
Positive
The DUT (battery)
To the + (positive) terminal
To the – (negative) terminal
DUT cable
Negative
Spring washer
Nut
PFX2532
Fasten the screws.
PFX2532