116 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
Condition Errors 65 to 92, Calibration Needed (Extended)
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:NEEDed sub-register. The
second column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 72 indicates that the “Input Attenuation not calibrated”
condition has been detected, and error 1072 indicates that failure has been
cleared.
This register is summarized as bit 8 of the STATus:QUEStionable:CALibration
register. See section
“Condition Errors 36 to 64, Calibration Needed or Failed”
Err#
Bit in
status
register
Message
Error or
Warning
More Information
3
System alignment required
65
0
unused
66
1
Align 9kHz–30MHz required
E
An EMI conducted frequency range
alignment is needed.
68
2
Align 30MHz–1GHz required
E
An EMI radiated frequency range
alignment is needed.
72
4
Input Attenuation not
calibrated
E
Corrected measurements have been
requested and the required RF
front-end setting of x dB has not been
calibrated.
74
5
unused
76
6
unused
78
7
unused
80
8
unused
82
9
unused
84
10
unused
86
11
unused
88
12
unused
90
13
unused
92
14
unused
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...