Keysight Technologies N9040B UXA Signal Analyzer Service Guide 417
Assembly Replacement Procedures
Card Cage Boards
A21 Wideband Analog I.F. Assembly (Option B2X and B5X only)
Removal
1.
To remove the A21 Wideband Analog I.F. assembly it is also necessary to
remove the top brace. Perform
in the
removal procedure.
2.
Refer to
. Locate the Wideband Analog I.F. assembly
(3)
.
3.
Refer to
. Remove ribbon cable W56 by first removing the
single screw
(1)
(0515-0372). Then unplug W56 from the interconnection
with W57.
Figure 15-67
Wideband I.F. Ribbon Cable
4.
Refer to
. Remove cables W34, W40, W41, W45, and W46.
Figure 15-68
Wideband I.F. Cables
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...