Keysight Technologies N9040B UXA Signal Analyzer Service Guide 451
Assembly Replacement Procedures
Fan Assembly
4. For Serial Number Prefix
≥
5605 and All Options 544 and 550
:
—
Refer to
. Remove the seven screws
(1)
(0515-0372)
and the six screws
(2)
(0515-1946) that attach the fan bracket
(3)
to the chassis.
Figure 15-99
Fan Bracket Removal
Serial Number Prefix
≥
5605 and All Options 544 and 550