Keysight Technologies N9040B UXA Signal Analyzer Service Guide 375
Assembly Replacement Procedures
RF Area - Options 508, 513, 526
Replacement
1.
. When replacing the coax switches, install them into
the bracket with the label facing as shown. Install SW1 (bottom switch)
and secure with the two screws
(1)
(0515-1992) and SW1 bracket
(2)
(E4410-00104) to the main switch bracket
(4)
. Torque to 6 inch-lbs.
Install SW2 (top switch) with the two screws
(1)
(0515-1992) and SW2
bracket
(3)
(N9040-00019). Torque to 6 inch-lbs.
Figure 15-21
Coax Switches Replacement - Options 508, 513, 526
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...