124 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
207
2
Burst not found;with
selected Time Slot
E
The selected timeslot does not contain the
expected burst.
209
3
Timing Error
E
209
3
Timing Error:No time
ref pilot burst
E
The pilot burst used for time reference is not
active.
211
4
Carrier(s) incorrect
or missing
E
In the PSA, this error was reported as one of
the following error numbers: 10165, 10173,
10178, 10419, 10421, 10535, 10560,
10642, 10648, 10650, 10960
213
5
Freq Out of Range
E
213
5
Freq Out of Range;
System input (IF)
E
One or more system input frequencies are
out of range. If using a frequency list, check
that all entries are valid for current
measurement mode.
213
5
Freq Out of Range;
External LO
E
One or more external LO frequencies are out
of range. Check that the LO frequency limits
are set correctly and check the entered
measurement frequencies and
measurement mode.
215
6
Sync Error
E
W-CDMA: Cannot sync DPCCH pilot.
Cannot synchronize measurement with
DPCCH pilot for Power Control
measurement, because the pilot signal
cannot be found. Make sure DPCCH is
present in the W-CDMA uplink signal, and
that the slot format and scramble code are
set correctly.
215
6
Sync Error;No pilot
burst
E
There is no Pilot burst detected.
215
6
Sync Error;Sync code
not found
E
Synchronization code is not found in the
measured time slot.
215
6
Sync Error;No freq
ref pilot burst
E
The pilot burst used for frequency reference
is not active.
215
6
Sync Error;Midamble
sync fail
E
Failed to find the uplink slot, which caused
the synchronization with the midamble to
fail.
215
6
Sync Error;Preamble
length zero
E
Burst type is “Data” or “Preamble” and the
measurement cannot find a Preamble
Err#
Bit in
Status
Register
Message
Error or
Warning
More Information
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...