398 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area - Options 544, 550
A20 YTO
Removal
1.
Refer to
. Remove cable W20 from the YTO
(2)
.
2.
Remove the ribbon cable W23 by pulling forward on the locking latch.
3.
Refer to
. From the bottom of the instrument, remove the two
screws
(2)
(0515-0372). The A20 YTO can now be removed from the
chassis.
Replacement
1.
Refer to
. Place the YTO into the chassis. Replace the two
screws
(2)
(0515-0372). Torque to 9 inch-pounds.
2.
. Replace the cable W20. Torque to 10 inch-pounds.
3.
Replace the ribbon cable W23. Ensure the locking latch is engaged.
4.
Replace the RF bracket, front panel, and instrument outer case as
described on
.
Item
Keysight Part Number
W20
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...