Keysight Technologies N9040B UXA Signal Analyzer Service Guide 445
Assembly Replacement Procedures
A4 CPU/A5 Solid State Drive
2.
Refer to
. Remove the four screws
(5)
(0515-1035) from the
disk drive tray that secure the SSD
(1)
.
Figure 15-93
Disk Drive Tray Screws
Replacement
1.
Refer to
. Place the new SSD into the tray assembly and
attach with the four screws
(5)
(0515-1035). Torque to 9 inch-pounds.
2.
Refer to
. Slide the SSD into the CPU assembly and push to
mate the connector. Secure the thumb screw to 9 inch-pounds.
Make sure not to use the 0515-0372 screw because for the PC6 processor it will cause the SSD
to interfere with the CPU memory card.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...