Keysight Technologies N9040B UXA Signal Analyzer Service Guide 123
Instrument Messages
Condition Messages
207
2
Burst Not Found
E
The burst signal cannot be detected
because of inappropriate parameter settings
or incorrect signal.
An in appropriate parameter setting could
cause the signal to be partially, rather than
fully, on the display, Burst Search Threshold
and/or Burst Search Length may need to be
adjusted.
An incorrect signal could have either
insufficient power, the rising or falling edges
cannot be detected, or the burst is less than
126 microseconds.
Carrier signal is not actually bursted.
W-CDMA: Either the signal being analyzed
has insufficient power, the rising or falling
edges cannot be detected, or the burst is
less than 126 microseconds.
W_CDMA: Cannot synchronize
measurement with PRACH channel for
Power Control measurement, because the
signal cannot be found. Make sure PRACH is
present in the W-CDMA uplink signal, and
that the preamble signature and scramble
code are set correctly.
GSM: Data was acquired but a GSM burst
was not found, with the timeslot mode
disabled.
NADC, PDC: A valid burst is not found when
the Device is MS.
1xEV-DO: Data was acquired but a 1xEV
burst was not found, with the timeslot mode
disabled.
Bluetooth
: The burst that has been found
does not correspond to the currently
selected
Bluetooth
packet type (the burst
length may be too short).
WLAN: The instrument cannot find a valid
WLAN burst. You may need to extend the
search length.
In the PSA, this error was reported as one of
the following error numbers: 10772, 13104,
10160, 10286, 10420, 10454, 10614,
10904, 10928, 13074, 10287
Err#
Bit in
Status
Register
Message
Error or
Warning
More Information
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...