Keysight Technologies N9040B UXA Signal Analyzer Service Guide 91
Instrument Messages
Event Messages
–221
Settings conflict;
Destination trace for Trace
Math cannot be a trace
operand
The resulting trace data (from doing a trace math function)
cannot be put into the any of the traces that are being used
by the math operation.
–221
Settings conflict; Downconv
only available when DUT is
Amplifier
SCPI only message. The System Downconverter can only be
set to ‘On’ when the DUT type is amplifier. Change DUT type
to Amplifier if the System Downconverter is required.
–221
Settings conflict; EDGE EVM
only supports EDGE TCH burst
type.
–221
Settings conflict;
Electronic attenuator is
disabled
You are using the mechanical attenuator, and have not
enabled the electronic attenuator. You cannot set the value
of the electronic attenuator because it automatically
sets/changes when enabled.
–221
Settings conflict;
Electronic attenuator is not
available above 3.6 GHz
The maximum frequency of the electronic attenuator is 3.6
GHz. This is because of switching capacitance.
–221
Settings conflict;
Electronic attenuator
unavailable in current state
–221
Settings conflict;
Electronic attenuator
unavailable with Preamp on
The internal preamp is on. Electronic attenuator cannot be
used while you are using the internal preamp.
–221
Settings Conflict; FAST
method can only be used
while Radio Std is W-CDMA
–221
Settings conflict; Feature
not available in this View
Some functionality is available in one View, but not in
another. (See the Views under the View/Display key.) This
error occurs if you send a SCPI command or push a gray-out
key that is not available in the current selected View.
–221
Settings conflict; Feature
not supported for selected
source
You have asked for a feature that the selected source is not
capable of.
–221
Settings conflict; Feature
not supported for this
measurement.
Some functionality is available in one measurement, but not
in another. (See the measurements under the Meas key.)
This error occurs if you send a SCPI command or push a
gray-out key that is not available in the current selected
measurement.
–221
Settings conflict; Feature
not supported for this model
number
This functionality is not a part of the instrument you are
using but may be found in other models in the X-Series.
Err#
Message
Verbose/Correction Information
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...