Keysight Technologies N9040B UXA Signal Analyzer Service Guide 179
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Verifying SW3 (Option LNP only)
Since there is a ribbon cable from the A15J802 to SW3, it is not feasible to use
a voltmeter or oscilloscope to measure the control voltages. For testing SW3,
the “click” test can be used.
Press
Mode Preset
on the analyzer. Press
FREQ
,
5 GHz
,
SPAN
,
1 MHz
. Verify
the optional preamplifier is off by pressing
AMPTD
,
Signal Path
and verify the
Internal Preamp Low Band toggle is in the Off position. The Internal Preamp
toggle must be Off in order to test SW3. Press
AMPTD
,
Signal Path
,
uW Path
Control
,
Low Noise Path Enable
on the analyzer. You should hear a distinct
“click” noise from SW3 when the low noise path is selected. Now press
uW
Path Control
and select
Standard Path
. SW3 will “click” again and return to
the standard RF path.
Another way of testing this switch is using an RF signal. This is described in the
RF Section Troubleshooting chapter, under the high band troubleshooting
section.
Verifying SW1 and SW2 (Option MPB)
SW1 and SW2 can be verified three different ways:
— “Click” Test
— R.F. Signal Tracing
“Click” Test
This test changes between two different path modes. When this is done a
distinct “click” noise from SW1 and SW2 can be heard. Since both switches
switch at the same time, it is difficult to decipher if one is switching and the
other is not.
Press
Mode
,
Preset
on the analyzer. Press
FREQ
,
5 GHz
,
SPAN
,
1 MHz
,
AMPTD
,
Signal Path
,
uW Path Control
. Standard Path is the default. Select
uW Preselector Bypass and you should hear a distinct “click” noise if the
control logic from the A15 Front End Control board is getting to SW1 and SW2.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...