100 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–200 to –299, Execution Errors
For –221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
–21x – Trigger errors
–22x – Parameter error
–23x – Data corrupt or stale (invalid data)
–24x – Hardware error
–25x – Mass storage error
–26x – Expression data error
–27x – Macro error
–28x – Program error (a downloaded program-related execution error)
–29x – Memory use error
Err#
Message
Verbose/Correction Information
–200
All ranges are off. Turn
on at least a range
There are no range turn on in scan table. You need to turn on at
least a range to initiate a scan.
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom.
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom.
–200
Execution Error
A program execution error has occurred. The exact problem
cannot be specifically identified.
–200
Execution error; Carrier
frequency outside device’s
transmit band
The entered channel/carrier frequency is not within the range of
your current mode setup selection of standard and device.
–200
Execution error; Invalid
GSM burst timing
A GSM-like burst was acquired, but its timing is not valid. Ensure
the correct Burst Type has been selected.
–200
Execution error; Invalid
IP address
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
–200
Execution error; Invalid
Marker Trace.
Cannot place markers on the reference trace, because the
reference trace is currently turned off or has no data.
–200
Execution Error; No peak
found.
No signal peak was found within the defined parameters of the
search.
–200
Execution error; No ranges
are defined. Activate a
range.
There are no active ranges in the range table. You will need to
activate at least one range.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...