Keysight Technologies N9040B UXA Signal Analyzer Service Guide 85
Instrument Messages
Event Messages
–800, Operation Complete Event
–700, Request Control Event
–600, User Request Event
–500, Power on Event
–400 to –499, Query Errors
Err#
Message
Verbose/Correction Information
–800
Operation complete
The instrument has completed all selected pending operations in
accordance with the IEEE 488.2, 12.5.2 synchronization protocol.
Err#
Message
Verbose/Correction Information
–700
Request control
The instrument requested to become the active IEEE 4881
controller-in-charge.
Err#
Message
Verbose/Correction Information
–600
User request
The instrument has detected the activation of a user request local
control.
Err#
Message
Verbose/Correction Information
–500
Power on
The instrument has detected an off to on transition in its power
supply.
Err#
Message
Verbose/Correction Information
–400
Query Error
There was a problem with a query command. The exact problem
cannot be specifically identified.
–410
Query INTERRUPTED
Some condition caused an INTERRUPTED query to occur. For
example, a query was followed by DAB or GET before a response
was completely sent.
–420
Query UNTERMINATED
Some condition caused an UNTERMINATED query to occur. For
example, the device was addressed to talk and an incomplete
program message was received.
–430
Query DEADLOCKED
Some condition caused a DEADLOCKED query to occur. For
example, both the input buffer and the output buffer are full and
the analyzer cannot continue. The analyzer automatically discards
output to correct the deadlock.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...