270 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Service and Diagnostics Menus
Service Key Descriptions
Alignments
The Alignments Menu allows accessing Diagnostic capabilities of Alignment,
and invoking alignments for individual subsystems.
Diagnostics
The Diagnostics menu contains items for controlling the operating behavior
of Alignment and the Alignment reporting capabilities.
Visible Align
Controls the state of Visible Align. Visible Align replaces the “Aligning
1 or X” messages with descriptive messages describing the
alignment piece begin executed.
Align Log…
(
This is
not
implemented in A.15.xx software
)
Invokes Notepad with the Alignment Log loaded. The log can be
viewed or saved to an external media or drive. Notepad can be closed
without a mouse or external keyboard by pressing ALT front-panel
key, then arrow down to highlight Exit, then press Enter.
This file can also be accessed directly on the instrument hard drive
as:
E:\AlignDataStorage\AlignmentHistory.txt
Align Log Mode
(
This is
not
implemented in A.15.xx software
)
The Alignment System places information about the last alignment
performed into a log. The log can be configured to operate in clear or
append mode. In clear mode the log only contains information on the
last alignment performed; the log is cleared with each alignment
performed. In append mode each new alignment appends to the log.
In append mode, care must be taken by the operator to not allow the
log to grow to a size so large as to inhibit system operation.
Immediate Actions
The Immediate Actions menu allows invoking individual components of the
internal alignments.
ADC
Immediately executes an alignment of the ADC subsystem. The
instrument will stop any measurement currently underway, perform
the alignment, then restart the measurement from the beginning.
A failure of ADC will set the Error Condition “Align ADC failed”. A
failure will not employ new ADC alignment data.
Successful completion of ADC will clear the Error Condition “Align
ADC failed”.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...