Keysight Technologies N9040B UXA Signal Analyzer Service Guide 469
Assembly Replacement Procedures
Front Frame Assembly
Front Frame Assembly
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
Refer to
. Disconnect the two cables
(1)
(just behind the
front frame) with a 5/16-inch wrench and a 1/4-inch wrench.
Figure 15-114
EXM Cable
Make sure any connectors on the front panel are removed.