220 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
CPU/Disk Drive Troubleshooting
A5 Disk Drive Troubleshooting
Disk Drive Recovery Process
The Instrument Image Recovery System is stored in a separate hidden disk
drive partition. It can be used to attempt to repair errors or restore the original
factory instrument system on the C: partition of the instrument disk drive.
Using the Instrument Image Recovery System
1.
Make sure the instrument is turned off
2.
Turn on the instrument
3.
After the "Keysight Technologies" screen is displayed the Windows Boot
Manager screen as shown in
will be displayed for 5 seconds
Figure 8-2
Windows Boot Manager
4.
Press the down arrow key to move the highlight to "Instrument Image
Recovery System", and press the
Enter
key.
5.
When the Instrument Image Recovery System has booted, follow the
on-screen instructions to recover the image of the C: partition.
6.
After exiting the Instrument Image Recovery System, the instrument may
reboot a few times.
7.
Update the instrument application software to the latest version by
downloading it from the following URL:
http://www.keysight.com/find/uxa_software
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...