Keysight Technologies N9040B UXA Signal Analyzer Service Guide 531
Functional Tests
Scale Fidelity
Table 17-8
Scale Fidelity Results
External Attenuator Setting
Minimum
(dB)
Marker Delta Value
(dB)
Maximum
(dB)
0
N/A
Reference
N/A
10
–11.0
–9.0
20
–21.0
–19.0
30
–31.0
–29.0
40
–41.0
–39.0
50
–51.0
–49.0