Keysight Technologies N9040B UXA Signal Analyzer Service Guide 441
Assembly Replacement Procedures
A6 Power Supply
Serial Number Prefix
≥
5605 and All Options 544 and 550
1.
If the instrument has Option B2X or B5X, the A22 and A23 Wideband
Digital I.F. boards must be removed. Refer to the
removal procedure.
2.
Refer to
. Remove the nine screws
(2)
(0515-1946) that
attach the A6 power supply
(1)
to the bracket. The power supply can now
be removed by sliding it out of the rear of the chassis.
Figure 15-90
Power Supply Removal
Serial Number Prefix
≥
5605 and All Options 544 and 550
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...