Keysight Technologies N9040B UXA Signal Analyzer Service Guide 107
Instrument Messages
Event Messages
–250
Mass storage error; Pk
Table must be on to save
Pk Table as Meas Results
You have to have a Peak Table on the screen before you can save
it. Turn on the Peak Table and try again.
–250
Mass storage error; Read
fault
The system cannot read from the specified device.
–250
Mass storage error;
Register <number> empty
Attempt to recall a register with nothing in it
–250
Mass storage error;
Sharing violation
The process cannot access the file because it is being used by
another process.
–250
Mass storage error;
Spectrogram must be on to
save as Meas Results
You have to have a Spectrogram on the screen before you can
save it. Turn on the Spectrogram and try again.
–250
Mass storage error; Too
many open files
The system cannot open the file.
–250
Mass storage error; Write
fault
The system cannot write to the specified device.
–252
Missing media
A legal command or query could not be executed because
missing media.
–253
Corrupt media
A removable media was found to be bad or incorrectly formatted.
Any existing data on the media may have been lost.
–254
Media full
A legal command/query could not be executed because the
media was full
–255
Directory full
A legal command or query could not be executed because media
directory was full.
–256
File name not found;
A legal command or query could not be executed because the file
name was not found in the specified location.
–257
File name error;
A legal command or query could not be executed because there
was an error with the file name on the device media. For example,
maybe you tried to copy to a duplicate file name.
–257
File name error; Allowable
extension is .csv
You are using the wrong type of file extension for the current
data/file type.
–257
File name error; Allowable
extension is .png
You are using the wrong type of file extension for the current
data/file type.
–257
File name error; Allowable
extension is .state
You are using the wrong type of file extension for the current
data/file type.
–257
File name error; Invalid
file name
The filename, directory name, or volume label syntax is incorrect.
–257
File name error; name too
long
Err#
Message
Verbose/Correction Information
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...