Keysight Technologies N9040B UXA Signal Analyzer Service Guide 515
Functional Tests
Second Harmonic Distortion (SHD)
9.
Press
Peak Search
. Enter the displayed value under the Measured Second
Harmonic Distortion (dBc) heading in
Table 17-3
Second Harmonic Distortion Results
Applied Frequency
(MHz)
Measured Second Harmonic Distortion
(dBc)
Mixer Level
(dBm)
Specification
(dBc)
40
–15
–60