[SOURce:]CURRent:PROTection:STATe ON|1|OFF|0
[SOURce:]CURRent:PROTection:STATe?
Enables or disables current protection.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Enable limit control operation: CURR:PROT:STATe OFF
ON (TRIP): Turns OUTPUT off and activates the alarm for overload conditions longer than three
seconds.
OFF (LIMIT CONTROL): Decreases output voltage to keep current at or below the limit when an
overload occurs.
To clear an overcurrent condition, remove the condition's cause and send
.
If overcurrent protection is enabled and the output goes into current limit, the output is disabled and
the Questionable Condition status register OCP bit is set.
The
command sets this parameter to ON.
3 SCPI Programming Reference
122
Keysight AC6800B Series Operating and Service Guide