USB Configuration
The Keysight IO Libraries are required to control the instrument through the USB interface.
Use a standard USB cable to connect the instrument to the computer. There are no configurable USB
parameters, but you can retrieve the USB connect string using the front panel.
Front Panel
SCPI Command
[Menu] > System > IO > USB
The USB connect string appears.
Not available
The instrument complies with USB Specification 2.0, USBTMC Specification 1.0, and USBTMC-
USB488 Specification 1.0. The maximum data rate is 12.5 Mbps, the vendor ID is 0x0957, and the
product ID values are shown in the table below.
Model
Product ID
AC6801B
0x1A02
AC6802B
0x1B02
AC6803B
0x1C02
AC6804B
0x1D02
GPIB Configuration
The GPIB interface requires the optional GPIB interface board and uses a standard IEEE-488 cable to
connect to the computer.
Each device on the GPIB (IEEE-488) interface must have a unique whole number address between 0
and 30 (default 5). Your computer’s GPIB interface card address must not conflict with any instrument
on the interface bus. To change the GPIB address:
Front Panel
SCPI Command
[Menu] > System > IO > GPIB
Use the numeric keypad to enter a value from 0 to 30. Then press [Enter].
Not available
This setting is nonvolatile; it will not be changed by power cycling or
.
LAN Configuration
The following sections describe the primary front-panel LAN configuration functions. There are no
equivalent SCPI commands.
To begin configuring the remote interface from the front panel, press [Menu] > System > IO > LAN.
1 Getting Started
42
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...