*PSC 0|1
*PSC?
Enables (1) or disables (0) the clearing of certain enable registers at power on. These include:
Questionable Data Register (STATus:QUEStionable:ENABle)
Standard Operation Register (STATus:OPERation:ENABle)
Status Byte Condition Register (*SRE)
Standard Event Enable Register (*ESE)
The *PSC command does not affect the clearing of the condition or event registers,
just the enable registers. For more information on the SCPI Status System, see
Status Subsystem Introduction.
Parameter
Typical Return
0|1, default 1
0 or 1
Disable power-on clearing of affected registers: *PSC 0
*RCL <
0 to 10
>
Recalls a saved instrument state.
Parameter
Typical Return
0 to 10
(none)
Recall state from location 1: *RCL 1
This command restores the instrument to a state that was previously stored with the *SAV command.
All instrument settings are recalled except: (1) trigger systems are set to the Idle state, (2) calibration
is disabled, (3) nonvolatile settings are not affected.
This command also aborts TRANsient and ACQuire operations.
Location 0 is recalled at power-on when the Output Power-On state is RCL0.
Stored instrument states are not affected by *RST.
*RST
Resets the instrument to default values. These settings are described in
.
Parameter
Typical Return
(none)
(none)
Reset the instrument: *RST
3 SCPI Programming Reference
136
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...