Field
Display
Description
Status
Off
Output is disabled
CV
Constant voltage mode
CLrms
Current limit mode
PL
Power limit mode
OV
Overvoltage protection; output OFF
OC
Overcurrent protection (rms/avg); output OFF
OCPK
Overcurrent protection (peak); output OFF
OP
Overpower protection; output OFF
OT
Overtemperature protection; output OFF
LV
Low voltage protection (AC input); output OFF
WDG
Watchdog timeout error; output OFF
SF
Sense fault protection; output OFF
Range
LOW
Selected range up to 155 VAC (219 VDC)
HIGH
Selected range up to 310 VAC (438 VDC)
AUTO
Autoranging selected (not available in EXT-AC or EXT-DC mode)
Interface Err
One or more messages in error queue
LAN (displayed)
LXI LAN "No fault"
LAN (blinking)
LXI LAN "Identify"
(LAN section blank) LXI LAN "Fault"
IO
Remote activity via the LAN connection
If AUTO is selected prior to entering either of the external programming modes, the range selected by
autoranging is retained until changed by the user. The range field is updated to reflect the actual
range when the external programming mode is first entered (if AUTO is selected).
2 User Information
70
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...