Voltage Ranges and Limits
The instrument has two voltage ranges: 155 V (low) and 310 V (high). It can also autorange between
the two ranges in AC or DC mode. The instrument will not switch voltage ranges (155 V, 310 V, or
AUTO), with the output on. If you attempt to do so, the output shuts off. Also, if you switch the range to
155 V when the voltage is set above 137.5 V the instrument sets the output voltage to 0 V.
Autoranging automatically switches to the 155 V or 310 V range according to the specified voltage. As
it does so, if the output is on, the instrument turns the output off for approximately 0.5 seconds and
turns it on again after the range switches.
The tables below indicate the output voltage setting ranges and maximum output current.
Voltage range Output Voltage Setting
AC Mode
DC Mode or AC+DC Mode
155 V
0.0 to 157.5 V
-222.5 to +222.5 V
310 V
0.0 to 315.0 V
-445.0 to +445.0 V
Voltage range Maximum Output Current
AC Mode DC Mode or AC+DC Mode
AC6801B
155 V
5 A
4 A
310 V
2.5 A
2 A
AC6802B
155 V
10 A
8 A
310 V
5 A
4 A
AC6803B
155 V
20 A
16 A
310 V
10 A
8 A
AC6804B
155 V
40 A
32 A
310 V
20 A
16 A
Keysight AC6800B Series Operating and Service Guide
57
1 Getting Started
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...