Step
Description
Front Panel
SCPI
1
Select the low voltage range.
[Menu] > System > Admin > Cal >
AC.
Then select the
Low
range.
CAL:VOLT 155
2
Select the first voltage
calibration point.
Select
Next
. The information field should
indicate:
Enter P1 measured data
CAL:LEV P1
*OPC?
3
Measure the output voltage
with the DMM and enter the
data.
Enter the measurement from the external
DMM in the
Measured Data
field. Press
[Select]
or
[Enter]
when done.
CAL:DATA <
data
>
4
Repeat steps 2 and 3 for the second, third, fourth, and fifth calibration points. Note that the SCPI com-
mand should use P2, P3, P4, and P5 in place of P1. Similarly, the front-panel
Enter P1 measured
data
message from step 2 will indicate P2, P3, P4, or P5.
5
Select
Next
or press
Back
to finish this part of the calibration.
AC Voltage High Range
Use the same DMM configuration as for the AC Voltage Low Range calibration.
Step
Description
Front Panel
SCPI
1
Select the high voltage
range.
[Menu] > System > Admin > Cal >
AC.
Then select the
High
range.
CAL:VOLT 310
2
Select the first voltage
calibration point.
Select
Next
. The information field should
indicate:
Enter P1 measured data
CAL:LEV P1
*OPC?
3
Measure the output voltage
with the DMM and enter the
data.
Enter the measurement from the external
DMM in the
Measured Data
field. Press
[Select]
or
[Enter]
when done.
CAL:DATA <
data
>
4
Repeat steps 2 and 3 for the second, third, fourth, and fifth calibration points. Note that the SCPI com-
mand should use P2, P3, P4, and P5 in place of P1. Similarly, the front-panel
Enter P1 measured
data
message from step 2 will indicate P2, P3, P4, or P5.
5
Select
Next
or press
Back
to finish this part of the calibration.
Current Low Range
Connect a precision shunt resistor to the output. The shunt resistor should be able to measure at least
120% of the output’s full-scale current. Connect the Keysight 3458A DMM across the shunt resistor.
Connect the Keysight 6060B electronic load to the AC source's output. Select the full-scale (FS), Low
Range current measurement value from the Current Table for the model being calibrated, 4 A for the
AC6801B.
Keysight AC6800B Series Operating and Service Guide
173
4 Calibration, Verification, and Service
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...