Standard Event Status Group
These registers are programmed by Common commands. The group consists of an Event and Enable
register. The Standard Event event register latches events relating to communication status. It is a
read-only register that is cleared when read. The Standard Event enable register functions similarly
to the enable registers of the Operation and Questionable status groups. See
for a
description of each register.
The event status register bits are set when certain events occur during instrument operation.
The register is controlled by the IEEE 488 commands *ESE, *ESE?, and *ESR?.
Bit Value Bit Name
Description
0
1
Operation Com-
plete (OPC)
All commands before and including *OPC have completed.
1
2
(not used)
0 is returned
2
4
Query Error (QYE)
The instrument tried to read an empty output buffer, a command was received before a previous
query was read, or the input and output buffers are full.
3
8
Device Dependent
Error (DDE)
A device-specific error occurred.
4
16 Execution Error
(EXE)
An execution error occurred.
. A valid SCPI command may not be executed cor-
rectly depending on the instrument's conditions.
5
32 Command Error
(CME)
A command syntax error occurred.
6
64 Reserved
0 is returned
7
128 Power ON (PON)
Power has been cycled since the last time the event register was read or cleared.
3 SCPI Programming Reference
104
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...