STATus:QUEStionable:CONDition?
Queries the
for the
group. This is a read-only register, which
holds the instrument's live (unlatched) operational status. Reading the Questionable Status
Condition register does not clear it.
Parameter
Typical Return
(none)
<
bit value
>
Read questionable status condition register: STAT:QUES:COND?
The value returned is the binary-weighted sum of all bits set in the register.For example, to enable bit
2 (value 4) and bit 4 (value 16), the corresponding decimal value would be 20 (4 + 16).
The condition register bits reflect the current condition. If a condition goes away, the corresponding bit
is cleared.
A
clears this register, other than those bits where the condition still exists after *RST.
STATus:QUEStionable:ENABle <
value
>
STATus:QUEStionable:ENABle?
Sets the value of the
for the
group. The enable register is a mask
for enabling specific bits from the Questionable Event register to set the QUES (questionable
summary) bit of the Status Byte register. The
command clears all enable register bits.
Parameter
Typical Return
Decimal sum of the bits in the register, default 0. For example, to enable bit 2 (value 4), bit 3 (value 8),
and bit 7 (value 128), use decimal sum 140 (4 + 8 + 128). Default 0.
<
bit value
>
Enable bits 2 and 4 in the questionable enable register: STAT:QUES1:ENAB 24
For example, to enable bit 2 (value 4) and bit 4 (value 16), the corresponding decimal value would be
20 (4 + 16).
A
does not clear the enable register, but does clear the
Keysight AC6800B Series Operating and Service Guide
149
3 SCPI Programming Reference
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...