Observe Environmental Conditions
Do not operate the instrument near flammable gases or fumes.
The AC6800B Series instruments are Overvoltage Category II instruments that should only be
operated in a controlled, indoor environment subject to the following restrictions:
Operating: 0 to 40 °C (32 to 104 °F), 20% to 80% relative humidity, noncondensing
Storage: –10 to 60 °C (14 to 140 °F), 90% or less relative humidity, noncondensing
Altitude: Up to 2000 m
Provide Adequate Air Flow
Do not block the air intake at the front of the instrument or the exhaust at the rear.
The dimensions of each model are shown below. Fans cool the instrument by drawing air through the
front and exhausting it out the back. Allow at least 8 inches (20 cm) of space at the front and back of
the unit for air circulation.
All dimensions are in mm.
AC6801B
Keysight AC6800B Series Operating and Service Guide
15
1 Getting Started