CALibrate:PASSword "<
password
>"
Sets the Admin password, which is used for calibration.
Parameter
Typical Return
an alphanumeric string of 4 to 15 characters (none)
Set the password to 12345: CAL:PASS "12345"
The password must be 4 to 15 characters long and may include any of these characters:
Characters
Description
A-Z
Upper case letters
a-z
Lower case letters
0-9
Numeric digits
+ - ( ) . , <space> Plus, minus, parentheses, period, comma, space
The
must be ON to execute this command.
This command increments the calibration count by 1.
The password is case sensitive.
To change the password: unsecure calibration memory with the old code, then set the new code.
If you enter an empty string as the new password and save it with CAL:SAVE once, no password will
be required to newly enter the calibration mode the next time.
This setting is non-volatile; it will not be changed by power cycling or
.
CALibrate:SAVE
Saves the calibration data and date, not the administration password, into nonvolatile memory. Do
this at the end of the calibration to avoid losing changes.
Parameter
Typical Return
(none)
(none)
Store calibration constants into nonvolatile memory: CAL:SAVE
The
must be ON to execute this command.
This command increments the calibration count by 1.
Keysight AC6800B Series Operating and Service Guide
119
3 SCPI Programming Reference
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...