Making Output Power Connections
Connecting the Remote Sense Wires
Preparation
Wire Requirements
For connecting the load, use noncombustible load wires rated to carry the maximum rated output
current.
Nominal Cross-Sectional Area (mm
2
) AWG Reference cross-sectional area (mm²) Allowable Current (A) (Ta=30 °C)
0.9
18
0.82
17
1.25
16
1.31
19
2
14
2.08
27
3.5
12
3.31
37
5.5
10
5.26
49
8
8
8.37
61
14
5
13.3
88
The values vary depending conditions such as the wire covering (insulator) and material (allowable
temperature) and whether they are multi-core cables. For cables other than those specified in the
table above, please consult with qualified personnel.
Output Terminal Cover
The output terminal is located at or near the center of the rear panel. The terminal cover location is
outlined in blue. Note that the sense connector on your unit my not be in the exact location as the one
shown in this figure.
Keysight AC6800B Series Operating and Service Guide
51
1 Getting Started
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...