Step Action
Normal Result
20
Record voltage reading of DMM.
21
Check test results
The difference between the DMM readings in Steps 18 and 20 are
within specified load effect limits.
22
Disable UUT. Open S1.
23
Turn on the UUT. Program output to RANGE 310 V,
VOLT 160, FREQ 500
Voltage at 160 V rms, Output current near zero
24
Record voltage reading of DMM.
25
Close S1. Turn on the UUT.
Voltage at 160 V rms, CV annunciator on, output current near:
2.5 A for AC6801B
5 A for AC6802B
10 A for AC6803B
20 A for AC6804B
26
Record voltage reading of DMM.
27
Check test results
The difference between the DMM readings in Steps 24 and 26 are
within specified load effect limits.
Keysight AC6800B Series Operating and Service Guide
181
4 Calibration, Verification, and Service