Retrofitting Options
4
Keysight 53210A/53220A/53230A Assembly Level Service Guide
207
To Retrofit the Internal Battery DC Power (Option 300)
Option 300 DC power input assembly parts
Preliminary steps
1
Remove the cover as follows:
2
To remove the counter handle, rotate the handle to the vertical (90 degree)
position and pull outward on the two side arms.
3
Using a T20 TORX screwdriver, remove the two T20 TORX screws securing the
rear panel bezel.
Item
Keysight part no.
Qty.
Battery charger PCB assembly
53200-66507
1
Battery
1420-0909
1
Y-Cable
53200-61602
1
Ribbon cable, 6-pin
53200-61603
1
Deck
53200-00601
1
Cover
53200-00602
1
TORX 10 screws
0515-0430
7
TORX 20 screws
0515-0433
2
CAUTION
Before proceeding with the installation, follow all the ESD steps as outlined
in the section titled
under the
at the beginning of
Chapter 4 of this service guide.
NOTE
Refer as necessary to the figures in the section titled
on page 185 of this guide.